Applications: TS-TA800 can be used to determine the moisture ash and volatile matter as well as calculation of fixed carbon and calorific value in coal coke calcined petrol coke and other solid biomass fuel etc. Dual Furnaces dual int...
Highlights:
High precision & accuracy.
①Two internal balances away from high temperature and heat radiation, no drift occurs.
②With automatic calibration function, each weighing is accurate and reliable.
③Simulate test environment and condition of drying oven and muffle furnace, exact in accordance with standard.
④The unique automatic cleaning function of low temperature furnace completely eliminates the influence of dust on result.
High efficiency
①Dual furnaces design, low-temperature furnace for moisture test and high- temperature furnace for ash and volatile matter test.
②Analysis for moisture, ash and volatile matter can be conducted simultaneously and consecutively.
③No blank sample during test. Full index of single batch samples (18 samples) can be finished in 150 min.
Patented technology
①Tailored volatile matter crucible with unique auto-sample loading mechanism avoided the operation of replacement or removal of the crucible cover under high temperature.
②Unique crucible clamp device avoid upset and drop of crucible.
③Unique crucible detection mechanism avoid mistake and missing of placing crucible.
④Unique sample tray positioning mechanism and drive mechanism ensure stable and reliable operation.
Human based design, Environment friendly and energy saving.
①Sample tray locates inside, real time data transmission. Weighing sample at room temperature, away from heat radiation.
②Special crucible collector discards residual automatically. No need to clamp crucible at high temperature. Easy and safe operation.
③Equipped with dust collector, environment friendly and clean.
④Compressed air can meet the requirement of gas supply, no oxygen and nitrogen required.
Intelligent control
①Easy to use Windows-based operating software, after finishing the sample weighing, the operator can be free for other tasks.
②Several proximate analyzers can be controlled by a single PC with Ethernet communication technology.
③Strong data processing ability, statistic report and print function, connection with network and balance are available by standard interface RS232.
Specification:
Description |
TS-TA800 |
Test index |
Moisture, ash and volatile matter |
Continuous test in batches |
Yes |
Standards |
GB/T 212-2008,ISO 11722, ISO 1171, ISO 562, ASTM-D 5142-2009 |
Precision |
0.02% RSD (1σ, 1g sample) |
Reproducibility |
Ash: 0.30% (<15.00%); 0.50% (15.00~30.00%); 0.70% (>30.00%)Volatile matter: 0.50% (<20.00%); 1.00% (20.00~40.00%); 1.50% (>40.00%) |
Cooling time required for batches of test |
No |
Furnace Cooling (forced air/N2) |
950 to 500°C<30 min |
Balance Resolution |
0.0001g |
Quantity of furnace |
2 |
Sample introduction and taking |
Automatic |
Quantity of sample/ Analysis time |
GB/ISO:18 / 150 min (all test index) ASTM:16 samples / 60 min~90 min (Moisture);16 samples/ 230 min (America Standard Ash), 10 samples / 7 min (ASTM Volatile matter 1) 10 samples / 36 min(ASTM Volatile matter 2) |
Sample weight |
0.5~1.1g |
Max. power |
4.5kw |
Furnace Temp |
Room temp. to 1000℃ |
Furnace Heating |
Resistance Wound |
Temperature Calibration |
Standard thermocouple |
Temp.control precision |
±2℃ (moisture) ±5℃ (ash, volatile matter) |
Gas Flow Rate |
GB:Moisture:Nitrogen,30~60 times/h Ash:Air 60~120 times/h ASTM:Moisture:Nitrogen or Air 2~4(Furnace Volume /min)volatile matter:Nitrogen 2~4(Furnace Volume /min) |
Power requirement |
220V(-15%~10%),50/60Hz |
Gas requirement |
Air/Nitrogen/Oxygen |
Gas Pressure |
Air 5 to 6 bar (75 to 90 psi) |
Minimum Gas Purity |
Air/Nitrogen/Oxygen 99.5%; dry, oil-free |
Sample weighing |
Under normal temp. |
Test Temperature |
Moisture 105℃,Ash 815℃,Volatile matter 920℃~950℃ Can be customized |
Humidity |
<80% |
Dimensions |
1055*747*656mm |
Weight |
211kg |
Operational Control |
Windows®-based software on external PC/Windows7/ WindowsXP |
Advance Calculation |
Yes |
Analysis Report |
Yes |
Statistics Report |
Yes |
System Monitor |
Yes |