Have any questions? robin@pro-trueview.com 0086-18007322239

You in this column



Bitumen Ductility Machine

Summary This instrument is designed and made as per national standard GB/T 4508-2010 Standard Test Method for Ductility of Bitumen and T0605-2011 Asphalt Ductility Test in JTG E20-2011 in the Industrial Standard of Peoples Republic of China...

Standard: ASTM -- D113 model: SYD-4508C-1 E-mail: robin@pro-trueview.com

Summary
  This instrument is designed and made as per national standard GB/T 4508-2010 Standard Test Method for Ductility of Bitumen and T0605-2011 Asphalt Ductility Test in JTG E20-2011 in the Industrial Standard of People’s Republic of China,Test Methods of Bitumen and Bituminous Mixture for Highway Engineering and ASTM D113 Standard Test Method for Ductility of Bituminous Materials. It is suitable to determine the distance to which it will elongate before breaking when two ends of a briquette specimen of the material are pulled apart at a specified speed and at a specified temperature,Unit:cm.

I.    Main technical characteristics
1.The the operation controller with independent structure design,large (800×480 dot matrix) color touch screen,two
    types:wire and wireless operation.
2.English version display screen with many kinds of flexible keys,inner real-time clock,mass memory.It can display and
    deal with many different detecting data such as temperature,pulse,optoelectronic input.The parameter setting,working
    status and data are all showed in English,more convenient to read.
3.It adopts the split design 50L water tank to independently cool by the water circulation,heat and control the constant
    temperature. 
4.It use the double water tanks design to make sure the constant temperature accuracy is at ±0.1℃ during the test
    time(30 mins) and there are no fluctuation in the water tank.It has the automatic reset after the test,and the system work
    again automatically.
5.With the advanced temperature control technology,short cooling time,good temperature control effect,uniform water
    bath temperature,At the standard 25℃ room temperature condition,the cooling period time is around 70mins from 25℃
    decreased to 5℃.
6.The stretch system adopts the external lead screw and double track location.The drive motor adopts the variable
    frequency motor,big torque,wide selection range,reliable and stable work.
7.With two work modes:automatic operation and manual,it makes the instrument operation and daily mantainance much
    convenient and easy.

II.    Main technical specifications
1.    Water bath temperature control range: 0~50℃
2.    Water bath temperature control accuracy: ±0.1℃
3.    Maximun measurement distance: ≤1500mm
4.    Measurement accuracy: 1mm
5.    Tensile speed: 50mm/min±2.5mm/min or 10mm/min±0.5mm/min
6.    Tensile motor power: 200W
7.    Heating power:3kW
8.    Refrigeration power: 2700 kilocalorie
9.    Power supply: Three phase 380V±10%、50HZ
10.    Environment temperature: -10℃~+35℃
11.    Relative humidity: ≤85%
12.    Boundary dimension: 2320mm×620mm×1030mm(L*W*H)
13.    Overall power consumption:less 4100W
 

ISYD-4508C-1 Bitumen Ductility Machine Accessories and Documents List

No. Name Unit Qty Memo
1 “SYD-4508C-1 Bitumen Ductility Machine Set 1  
2 Control cabnet Set 1  
3 Remote controller Piece 1  
4 Ductility mould base Pair 3 In same box
5 Ductility test module Pair 3
6 Thermometer Piece 2  
7 Voltage stabilization power line Piece 1  
8 Test module plate Piece 1  
9 Operation manual Piece 1  
10 Quality certificate Piece 1  
11 Product Warranty Card Piece 1  


IISYD-4508C-1 Bitumen Ductility Machine's Reagent and Auxiliary Equipments 


No. Name Specification Quantity Memo
1 Glycerol 500ml 1 bottle  
2 Bitumen solvent 2L 2L/bucket Kerosene
3 Talcum powder   500 g  
4 Scraper   1 piece  
5 Hairbrush 2cm width 2 pieces  
6 Electric furnace 1000W 1 piece  


Notes:
1The reagent and equipments for installing the instrument are supplied by end-users themselves.
2The usagemanagement and saving of poisonous and harmful reagentpls operate according to the relative principle.
3PLS read the operation manual and industry standard before using the instrument.