Summary 1.The instrument is designed and made as per the standards GB/T4509 Standard Test Method for Penetration of Bitumen T0604 Bitumen Penetration Test in the standard JTJ052-2000 Specifications and Test Methods of Bitumen and Bituminou...
Summary
1.The instrument is designed and made as per the standards GB/T4509 Standard Test Method for Penetration of
Bitumen, T0604 Bitumen Penetration Test in the standard JTJ052-2000 Specifications and Test Methods of Bitumen
and Bituminous Mixtures for Highway Engineering and ASTM D5.
2.The instrument is suitable to determine the penetration of highway bitumen, modified asphalt, as well as liquid
petroleum asphalt or residue of emulsified asphalt after evaporation.
3.The penetration of bituminous materials is expressed as the distance that a 100 g±0.05 g standard needle vertically
penetrates into the sample without measurable fraction in 5 s at a constant temperature bath of 25±0.1 ℃. It is used for
evaluating the consistency of asphalt. The larger the penetration is; the thinner and softer the asphalt is.
4.The instrument can also be used for inspecting industry materials such as solid granule, powder, colloid, as well as
food materials such as cream, glycine, etc. It has been widely used in the field of food industry, highway and traffic
departments and other departments.
I.Main technical specifications
1.Measurement range: 0 penetration~600 penetrations;
2.Resolution: 0.1 penetration;
3.Timing range: 0s~60 s, and the error is less than ±0.1 s.
4.Heating power: 200 W;
5.Temperature controlling accuracy: 25±0.1 ℃ (Note: the ambient temperature should not be higher than 20 ℃);
6.Data saved: 30 groups;
7.Constant temperature bath: harden glass container
a.Bath volume:5.5L
b.Temperature range: (5.0~100.0)℃
c.Temperature accuracy: ±0.1℃
d.Total consumption:less 1800W
e.Outline dimension: 420mm×380mm×900mm(L*W*H)
f:Power supply: AC(220±10%)V,50Hz
8.Stirrer: a magnetic stirrer;
9.Standard needle: 2.5±0.05 g and it conforms to the requirements of GB/T4509 and T0604.
10.Shifting holder: It has two ways to adjust the height of standard needle, so it is convenient to let the needle tip just
touch the sample surface.
11.Power supply: AC 220 V±10%, 50 Hz.
12.Outline dimension: 280mm×350mm×700mm(L*W*H)
No. | Name | Unit | Qty | Memo | ||
1 | SYD-2801I Automatic Penetrometer | Set | 1 | |||
2 | Low temperature constant temperature bath | Set | 1 | |||
3 | Standard needle(50㎜,2.5g±0.05g) | Piece | 3 | |||
4 | Weight (50g) | Piece | 1 | |||
5 | Asphalt sample container | Φ55×H35 (small sample container) | Piece | 3 | ||
Φ70×H45 (big sample container) | Piece | 3 | Optional | |||
Φ55×H70(special sample container) | Piece | 3 | Optional | |||
6 | Mercury thermometer (0~50 ℃ scale division is 0.1 ℃) | Piece | 1 | |||
7 | W-line | Piece | 1 | |||
8 | Power line(250V6A) | Piece | 1 | |||
9 | Triangle holder | Piece | 1 | |||
10 | Fuse(3A、Ф5×20) | Piece | 3 |
Used for penetration Instrument |
||
11 | Stylus printer | Piece | 1 | |||
12 | Regulated power supply (input 220V/50Hz output 5V/3A) | Piece | 1 | |||
13 | 25-pin data cable | Piece | 1 | |||
14 | Print paper (suitable for the above stylus printer) | Roll | 1 | Except installed in the instrument | ||
15 | Operation manual | Piece | 1 | |||
16 | Quality certificate | Piece | 1 | |||
17 | Product Warranty Card | Piece | 1 | |||
18 | Software CD | Piece | 1 |
No. | Name | Specification | Quantity | Memo |
1 | Glycerol | 500ml | 1 bottle | |
2 | Bitumen solvent(Kerosene) | 1L | 2L/bucket |
Bitumen test |
3 | Talcum powder | 500 g | ||
4 | Scraper | 1 piece | ||
5 | Hairbrush | 2cm width | 2 pieces | |
6 | Electric furnace | 1000W | 1 piece |
Bitumen test
|