Summary
This instrument is made as per standards T 0620-2000 and ASTM D2171 Dynamic Viscosity Test for Bitumen (Vacuum Capillary Method) in JTG E20-2011 Specification and Test Methods for Bitumen and Bituminous Mixture for Highway Engineering and SH/T0557-1998 Test method for Viscosity of Petroleum Asphalt (Capillary Method) . It is used to determine the dynamic viscosity of sticky petroleum asphalts by vacuum capillary viscometer.It is also suitable to determine other materials which viscosity range is 0.0036~20.000Pa.s.
I. Main technical features
1. It has a data base for coefficients of capillary viscometers. It can save 10 groups of coefficients at most. Operator can
callout the data whenever need.
2. It adopts high accuracy pressure sensor. The vacuum degree can be kept at 300mmHg±0.5mmHg. It can be shown on
the screen clearly.
3. The system will ignore the results which timing is less than 60s and enter next measurement step. The system will
calculate the viscosity automatically for the results which timing is beyond 60s.
4. Operator can choose and set automatic calculation of average value and bias percentage for 2~4 parallel tests at
will.The calculation can be done for several times to remove the unsatisfied results for operator. The test report can be
printed.
II. Main technical specifications
1. Temperature control mode: Point controlled by digital temperature controller.
2. Temperature control arrange: 0.00℃~100.00℃
3. Temperature control accuracy: ±0.01℃
4. Pressure range: 300mmHg±0.5mmHg
5. Timing mode: 4 built-in timer. Can do timing for 4 capillary viscometers
6. Timing range: 0.0s~99999.9s(27.7h); Bias ≤0.02%
7. Measurement range: About 18 Pa.s~580000Pa.s
8. Test samples: 4 samples
9. Operation interface: Touch-type colored LCD
10. Power supply: AC(220±10%)V , 50Hz
11. Ambient temperature: 5℃~50℃
12. Relative humidity: ≤85%
13. Overall dimension: 590mm×430mm×630mm
14. Net weight: 32.5kg(no water in bath)
15. Maximum power consumption: 1800W
III. Optional part:
1. Capillary viscometer washer:
SYD-0620-3 Capillary Viscometer Washing Apparatus
ISYD-0620B Bitumen Dynamic Viscometer Accessories and Documents List
No. |
Name |
Unit |
Qty |
Memo |
1 |
SYD-0620B Bitumen Dynamic Viscometer |
Set |
1 |
|
2 |
Vacuum capillary viscometer type: 50100200
Corresponding radius(mm):0.250.51.0 |
Piece |
Each 1 |
|
3 |
Mecury thermometer(50℃~100℃;scale mark 0.1℃) |
Piece |
1 |
|
4 |
Thermometer silica gel plug(1#) |
Piece |
1 |
|
5 |
Capillary clip |
Piece |
4 |
|
6 |
Stopping ring of thermometer(1#) |
Piece |
4 |
|
7 |
Bath cover |
Piece |
4 |
|
8 |
Fuse 10A(Ф5×20) |
Piece |
1 |
|
9 |
Silicone tube(External diameter Ф12×thickness2) |
Meter |
1 |
|
10 |
Operation manual |
Meter |
1 |
|
11 |
Temperature control device usage manual |
Meter |
1 |
|
12 |
Quality certificate |
Meter |
1 |
|
13 |
Repair guarantee |
Meter |
1 |
|
IISYD-0620B Bitumen Dynamic Viscometer's Reagent and Auxiliary Equipments
No. |
Name |
Specification |
Quantity |
Memo |
1 |
Trichloro ethylene |
Chemical pure 500ml |
5 bottles |
|
2 |
Distilled water |
Three grade |
20 litrs |
|
Notes:
1The reagent and equipments for installing the instrument are supplied by end-users themselves.
2The usagemanagement and saving of poisonous and harmful reagentpls operate according to the relative principle.
3PLS read the operation manual and industry standard before using the instrument.