Desc
Description:
1) Big LCD.
Making the electronic control part and column oven separately, to ensure the liquor and electronic control part is separated for convenient daily maintenance.
Using big LCD, which can display pressure, conductivity, output range, temperature compensation coefficient and pressure upper set at the same time. Displaying parameter in full screen, parameter has function of automatic memory, making customers easy to control the touch keys.
2) Having two way detection(normal detection and high precision detection)
3) Using import high-pressure dual piston plunger pump, which has a wide current velocity, and can bear high pressure(≤4200psi), acid-base and 100% organic solvent.
The current velocity of plunger pump is 0.001-10mL/min, increment is 0.001mL/min, can coordinate with different cation and anion for analysis.
Having articular infrared photoelectric sensor gas-liquid separator, which can avoid error and missing report.
4) PEEK flow path system
Using high-pressure six-way valve, having function of automatic signal acquisition and constant flow.
Full automatic gas removal technology, to avoid the analysis interference from dissolved gas in mobile phase.
5) Constant temperature system column, advanced digital temperature controlled conductivity detector.
Build-in two sets heating way, can make speedy heating-up and precision temperature control for the column oven.
It has high sensitivity, high stability, and precision detection with advanced digital temperature control conductivity detector.
6) Exoteric system design and compatibility
It can be compatible with other factory’s analysis column or suppressor besides cation and anion of IOC-600. It’s convenient for customers to improve the analysis way.
Can compatible with varies high capacity cation and anion column, to make analysis for the common anion and cation, DBPs, organic acid and so on.
Build-in air jack, which can run automatic injector.
Safety running with UPS stabilized power supply
7) Using Rheodyne 7725I Injector, to ensure precision sample injection. DS-PLUS technique can up to PPB level analysis.
7725I can use 10mL quantitative rods.
DS-PLUS technique can make the sample injection up to 1mL without effecting on the separated and quantitative water negative peak. It can up to PPB level analysis without provide beneficiation installation.
Particular imitating amplifying technique, advanced screening technique and precision temperature control design not only eliminate the negative effect of polarization and electric double layer, but also improve stability and accuracy for conductivity signal.
8) Advanced Suppressor and One-year guarantee
Using sixth generation suppressor technique can solve the problem of water negative peak drastically. High cost sodium carbonate liquor make chromatography be more stable and precision.
Continuous automatic renewable membrane suppressor, don’t need extra acid and can operate continuously.
One-year guarantee.
9) Anti-control function software
Chinese/English display. Having patent for multichannel data acquisition.
It has anti-control function for the processing system. Customer can choose hand control or anti-control. Can compatible with high-low speed and multi path running, also can make the anion and cation quickly detected to save time.
Specifications:
Ion Pump | |
Type | Imported High-Pressure Dual Piston Series Reciprocating Plunger Pump |
Maximum Pressure | 42Mpa |
Flow Rate Range | 0.001-9.999mL/min |
Pressure Precision | ≤0.1Mpa |
Flow Rate Precision | RSD<0.1% |
Flow Rate Repeatability | RSD≤0.1% |
Flow Path System | |
Full Plastic Flow Path | PEEK Material |
Six-way Valve | American Rheodyne High-Pressure Six-way Valve,can bear pressure 7000psi |
Constant Temperature Conductivity Detector | |
Type | The Fifth-Level Ring Passivity Electrode, 316 Stainless Steel Electrode Shield Conductivity Cell |
Detection Way | Single Conductivity Detection |
Cell Volume | ≤0.9 μL |
Resolution Ratio | ≤0.0020ns |
Detection Range | 0~35000 μS/cm(10 shifts for choosing) |
Output Voltage | ±6000mV |
Baseline Noise | <0.5% FS |
Baseline Drift | ≤±0.5% FS |
Temperature Control Range | Room Temperature +5℃~60℃ |
Temperature Precision | ±0.01℃ |
Temperature Compensation | 1.7%/℃ |
Limit Operation Pressure | 5.0MPa(750psi) |
Ion Detection | |
Anion Analysis | F-, Cl-, Br-, I-, NO2- , PO43- , NO3- , SO42- , ClO2- , BrO3- , ClO3- and so on, organic acid analysis |
Cation Analysis | Li+, Na+, NH4+, K+, Mg2+, Ca2+, Sr2+, Ba2+and so on, and betaine, choline chloride, organic amine analysis |
Analysis Repeatability | ≤1.0 %(calculated with SO42 –、Na+) |
Linear Range | ≥103 |
Limit Detection | ≤0.5 ppb(calculated with Cl-) ≤5 ppb(calculated with Na+) |
Quantitative Repeatability | RSD≤0.2% |
Sample Residual | ≤0.01% |
Gradient Error |
≤0.5% |