Automatic Chemistry Analyzer

Main features: Description: ●Bench top, Discrete, random access, fully automated; ●Constant speed 300 tests per hour; ●Unique Self-lubricating lead rail, accurate positioning and low noise ●Refrigerated reagent compartment with 45...

Standard: model: Bio-530 E-mail: robin@pro-trueview.com

Main features:

Description:

      ●Bench top, Discrete, random access, fully automated;
      ●Constant speed 300 tests per hour;
      ●Unique Self-lubricating lead rail, accurate positioning and low noise
      ●Refrigerated reagent compartment with 45 reagent positions;
      ●120 reusable cuvettes with 7-step auto wash;
      ●Automatic probe cleaning, liquid level detection & collision protection
      ●Reversed optic system with 8 wavelengths: 340~700nm (2 optional)
      ●Automatic dilution for abnormal sample
      ●External bar code reader (optional)
      ●Bi-directional LIS interface
      ●Multi-language available;

Technical Specifications:  (Enclosed Catalog)

System function:

Throughput: 300T/H( Single /double reagent)
Wavelength: 300nm-800nm
Source: 12V / 20W halogen lamps
Resolution: 0.0001A

Automatic biochemical analyzer Bio-530 loading system

Reagent needle: needle independent agents, plus the amount of reagent 10-500ul, 0.1ul step
Sample Needle: Independent sample needle, sample size 2-100ul, 0.1ul step

Automatic biochemical analyzer Bio-530 response system:

Reaction plate: using solid heating temperature control device to ensure a constant temperature
Stirring rod: independent high-precision polishing stirring rod, mix well, no carryover
Cleaning: 120 cuvettes, automatic cleaning

Automatic biochemical analyzer Bio-530 system software

Analysis Method: End-point, fixed time, kinetic method, supports single / dual wavelength
Operating system: Chinese/English user interface, windows 7/10/XP  system
A variety of single-point calibration method is linear, more linear and nonlinear: calibration method
QC rules: westgard more rules, L-J rule