Have any questions? robin@pro-trueview.com 0086-18007322239

You in this column



Pensky-Martens Closed-Cup Flash Point Tester

Summary This instrument is designed and manufactured as per standards GB/T 261-2008 Determination of Flash Point - Pensky-Martens Closed Cup Method and ASTM D93 Standard Test Method for Flash Point by Pensky-Martens Closed Cup Tester. It is...

Standard: ASTM -- D93 model: SYD-261(1991 Standard) E-mail: robin@pro-trueview.com

Summary
    This instrument is designed and manufactured as per standards GB/T 261-2008 Determination of Flash Point - Pensky-Martens Closed Cup Method and ASTM D93 Standard Test Method for Flash Point by Pensky-Martens Closed Cup Tester. It is used to make determination of the closed cup flash point of the petroleum products.
 
I. Main technical features
1. This instrument is designed as per GB/T261 and ASTM D93.
2. The heating power is continuously adjustable. The power is shown visually by voltmeter.The temperature control mode
     is advanced and reasonable.
3. The structure is designed small and exquisite.The stainless-steel table board is beautiful and easy to clean.
4. Operation is easy.Test results are precise.It has been chosen as the instrument to test the closed cup flash point by
     many metering and detecting institutes.
 
II. Main technical specifications
1. Power supply:AC (220±10%)V, 50Hz.
2. Heating device::The heating power is continuously adjustable from 0 W to 600W.
3. Heating rate::(1~12)℃/min;Controllable and adjustable.
4. Stirring rate: Procedure A: (90~120)RPM, Procedure B: (250±10)RPM
5. Oil cup:(1) Inner diameter: 50.7mm~50.8mm.
                  (2) Depth: 55.7mm~56.0mm 
                  (3) The scribed line depth of capacity of testing oil: 33.9mm~34.3mm.
                  (4) Capacity of testing oil:about 70ml
6. Igniting device:(1) Igniting source: gas(or other civilian fuels,the same below)
(2) Electric ignition. Gas flame diameter:3.2mm~4.8mm
7. Thermometers:Mercury-in-glass thermometer.Specifications are as below:
                              (1)Scale -5℃~110℃, division 0.5℃。
                              (2)Scale 20℃~150℃, division 1℃。
                              (3)Scale 90℃~370℃, division 2℃。
8. Ambient temperature:≤ 35℃
9. Relative humidity:≤ 85%
10.Maximum power consumption:650W

ISYD-261 Pensky-Martens Closed-Cup Flash Point Tester Accessories List


No. Name Unit Qty Memo
1 SYD-261  Pensky-Martens Closed-Cup Flash Point Test Set 1  
2 Thermometer -30~170℃ Piece 1  
3 Thermometer 100~300℃ Piece 1  
4 Driving flexible shaft Piece 1  
5 Latex tube φ3  250mm Piece 3 Including spare part 1 piece
6 Fuse  5A(Ф5×20) Piece 2  
7 Operation manual Piece 1  
8 Quality certificate Piece 1  
9 Products warranty certificate Piece 1  

IISYD-261 Pensky-Martens Closed-Cup Flash Point Tester(1991) Auxiliary Equipments


No. Name Specification Quantity Memo
1 Fuel gas Standard fuel gas steel bottle 1 bottle Liquefied gas bottle or pipeline gas
2 Fuel gas professional reduce valve   1 piece  
3 Fuel gas flexible pipe Normal 1.2~2 meters  
Notes:
1、The reagent and equipments for installing the instrument are supplied by end-users themselves.
2、The usagemanagement and saving of poisonous and harmful reagentpls operate according to the relative principle.
3、PLS read the operation manual and industry standard before using the instrument.