Have any questions? robin@pro-trueview.com 0086-18007322239

You in this column



Ultraviolet Fluorescence Sulfur,Nitrogen-in-Oil Analyzer

Summary The instrument is designed and manufactured according to the standards SH/T 0689-2000 Standard test method for determination of total sulfur in light hydrocarbons,motor fuels and oils by ultraviolet fluorescence,SH/T 0657-2007 S...

Standard: model: SYD-0689N E-mail: robin@pro-trueview.com

Summary 
  The instrument is designed and manufactured according to the standards SH/T 0689-2000 Standard test method for determination of total sulfur in light hydrocarbons,motor fuels and oils by ultraviolet fluorescence,SH/T 0657-2007 Standard test method for trace nitrogen in liquid petroleum hydrocarbons by syringe/inlet oxidative combustion and chemiluminescence detection and ASTM D 5453 -2006 standard.It is widely used in petroleum,chemical industry,electric power,coal,food,environment proection and others.At present,SYD-0689N is the advanced instrument to analyze the sulfur,nitrogen in China.

I.Main technical characteristics
1.    The instrument can detect the both sulfur and nitrogen,also could test the solid sample or liquid sample.
2.    With the desktop computer,Windows system,man-machine direct conversation platform,It can set all prameters and
       make the test by the keyboard operation.
3.    It adopts the control technology of the computer to automatically collect,process,store and print all data.with the wide  
       linearity range of measurement,strong capacity of resisting disturbance,easy operation and reliable test results.
4.    With the imported fluorescence excitation source,membrane desiccator,optical filter,metal encapsulated
       photomultiplier,It has the features of the high sensitivity,fast,stability,high precision,good consistency.
5.    With the advanced temperature control system,fast and steadily warming rate,long service life,the temperature control
       accuracy ±3℃ during the temperature range,It can meet the test requirements.  

II. Main technical specifications
1.    Determination method: Ultraviolet fluorescence method to detect sulfur,chemiluminescence to detect nitrogen
2.    Sample types:Liquid,solid and gas
3.    Sample injection quantity:Solid:1-20mg; Liquid:5-20μL; Gas:1-25mL
4.    Measurement range:  (0.2~8000)mg/L(High concentration can be diluted)
5.    Measurement accuracy:
Concentration values (ppm) Injection quantity(µL)  RSD(%)
0.2 20 25
5 10 10
50 10 5
100 10 3
5000 10 3
6.    Temperature control range: Ambient to 1050℃,accuracy: ±3℃
7.    Air supply requirements:High purity argon:above 99.995% ; 

                                                High purity oxygen:above 99.9%                      
Note:The water content of the gas must be less than 5ppm
8.    Environment temperature:Less 30℃(The laboratory with air condition)
9.    Relative humidity: ≤85%
10.    Power supply: AC220V±22V,50Hz±0.5Hz,more than 2000 W
11.    Dimension:Host:305(L)×460(W)×440(H)mm;                                                
        Temp controller:550(L)×460(W)×440(H)mm,No computer
12.    Net weight:Host::15kg, Temp controller:30kg
 

I,SYD-0689N Ultraviolet Fluorescence Sulfur,Nitrogen-in-Oil Analyzer Accessories and Documents 


No. Name Unit Qty Memo
1 Nitrogen analyzer host Set 1  
2 Sulfur analyzer host Set 1  
3 Filter Piece 1 Installed on reaction chamber
4 Pyrolyzer Set 1  
5 Silicone rubber pad Piece 20  
6 Silicone rubber tube Centimeter 20  
7 Rubber tube Meter 10  
8 Thermocouple Piece 1  
9 Power line Piece 3  
10 Communication line Piece 2  
11 Quartz tube Piece 1  
12 Injection needle Piece 4  
13 Membrane type dryer Piece 1  
14 Ozone generator Piece 1 Installed in machine
15 Photomultiplier Piece 1 Installed in machine
16 Ozone rid device Piece 1  
17 Ultravoilet lamp Piece 1  
18 Sulfur standard sample Piece 30  
19 Nitrogen standard sample Piece 30  
20 Special clamp for quartz cracking tube Piece 2  
21 Lenovo computer Set 1  
22 Laser printer Set 1  
23 Nitrogen reaction chamber Piece 1 Installed in machine
24 Sulfur reaction chamber Piece 1  
25 High temperature quartz cotton Packet 1  
26 Ground wire Piece 1  
27 Cylinder connector Piece 3  
28 Quartz tube bend Piece 2  
29 Liquid sampler injector Set 1  
30 Tweezers Pair 1  
31 O3 control connection line Piece 1  
32 Operation manual Piece 1  
33 Quality certificate Piece 1  
34 Product Warranty Card Piece 1  
 
II,SYD-0689N Ultraviolet Fluorescence Sulfur,Nitrogen-in-Oil Analyzer's Reagent and Auxiliary Equipments 

No. Name Specification Quantity Memo
1 High purity argon Purity above 99.95% 1 bottle With the reducing valve
2 High purity oxigen Purity above 99.95% 1 bottle With the reducing valve


Notes:
1、The reagent and equipments for installing the instrument are supplied by end-users themselves.
2、The usage,management and saving of poisonous and harmful reagent,pls operate according to the relative principle.
3、PLS read the operation manual and industry standard before using the instrument.