Summary
This instrument uses micro-coulometry technology computer to control the micro-coulomb titration. It’s the newest product, having reliable result, easy to operate, stable performance, easy to install to and so on. It is used to determine the trace sulfur of petro-chemical products. It is widely used in petroleum, chemical industry, and scientific research institutions. Standards: SH/T0253 Petroleum products-Determination of total sulfur content (Coulometry), SH/T0222, GB/T 11061.4-2010, GB/T 6324.4-1986, GB/T 12688.6-1990, GB/T 11141-1989, SY/T7508-1997, ASTM D3120, ASTM D3246.
I. Main technical features
1. Micro-coulometry technology. PC equipped. Windows OS.Easy to operate.
2. Oxidation style. It determines the sulfur or chlorine in petrochemicals as per Faraday's Law.
3. Wide application scope and good adaptation. It can be used to determine the sulfur or chlorine in liquid, gas or solid
materials.
4. Self-developed operation and test software. It will complete the data collecting, processing, saving and printing
automatically.
5. It needs less sample. Only 10μl for each test. The testing time is short. Only 1-2 minutes for each sample.
II. Technical specifications
1. Current: The maximum is ±2mA
2. Output voltage of amplifier: The maximum is ±30V
3. Bias voltage range: (0~500)mv, adjustable
4. Measuring range: 0.1mg/L~10000mg/L(dilutable for high concentration)
5. Repeatability error: ⑴ ≤50% when sample concentration <1.0mg/L
⑵ ≤10% when 1.0mg/L≤sample concentration ≤10mg/L
⑶ ≤5% when sample concentration >10mg/L
6. Temperature control: Ambient to 1000℃, ±1℃
7. Power supply: AC 220V±10V,50Hz±0.5Hz
8. Maximum power consumption: 3000W
9. Ambient temperature: (10~40)℃。
10. Relative humidity: ≤85%
11. Dimension: 700 mm×480 mm×540mm(PC is not included)
12. Net weight: 46kg(PC is not included)
I,SYD-0253A Coulometric Chlorine Analyzer Accessories and Documents List
No. |
Name |
Unit |
Qty |
Memo |
1 |
SYD-0253A Coulometric Sulfur Analyzer |
Set |
1 |
|
2 |
Liquid sampler injector |
Set |
1 |
|
3 |
Microsyringe |
Piece |
2 |
|
4 |
Standard sample |
Piece |
10 |
|
5 |
Silicone rubber pad |
Piece |
10 |
|
6 |
Silicone rubber tube |
Centimeter |
20 |
|
7 |
Pneumatic tube(Ф4) |
Meter |
10 |
|
8 |
Thermocouple |
Piece |
2 |
|
9 |
Power line |
Piece |
2 |
|
10 |
Communication cable |
Piece |
1 |
|
11 |
Quartz tube |
Piece |
1 |
|
12 |
Electrolytic cell |
Piece |
1 |
|
13 |
Quartz cracking tube exclusive clip |
Piece |
1 |
|
14 |
Brand computer |
Piece |
1 |
|
15 |
Printer |
Piece |
1 |
|
16 |
Stirrer |
Piece |
1 |
|
17 |
Operation manual |
Piece |
1 |
|
18 |
Quality certificate |
Piece |
1 |
|
19 |
Product Warranty Card |
Piece |
1 |
|
II,SYD-0253A Coulometric Chlorine Analyzer's Reagent and Auxiliary Equipments
No. |
Name |
Specification |
Quantity |
Memo |
1 |
Oxygen |
Purity 99.5% |
1 bottle |
With pressure regulating valve |
2 |
Nitrogen |
Purity 99.5% |
1 bottle |
With pressure regulating valve |
3 |
Glacial acetic acid(HAC) |
Excelent pure |
2 boxes |
|
4 |
Purified water |
|
|
Deionized water or redistilled water |
Notes:
1、The reagent and equipments for installing the instrument are supplied by end-users themselves.
2、The usage,management and saving of poisonous and harmful reagent,pls operate according to the relative principle.
3、PLS read the operation manual and industry standard before using the instrument.