Summary This instrument uses micro-coulometry technology computer to control the micro-coulomb titration. Its the newest product, having reliable result, easy to operate, stable performance, easy to install to and so on. It is used to deter...
Summary
This instrument uses micro-coulometry technology computer to control the micro-coulomb titration. It’s the newest product, having reliable result, easy to operate, stable performance, easy to install to and so on. It is used to determine the trace chlorine of petro-chemical products. It is widely used in petroleum, chemical industry, and scientific research institutions.Standards: SH/T 1757-2006, GB/T 18612-2011, ASTM D5808-09a, ASTM D5194-06.
I. Main technical features
1. Micro-coulometry technology. PC equipped. Windows OS.Easy to operate.
2. Oxidation style. It determines the sulfur or chlorine in petrochemicals as per Faraday's Law.
3. Wide application scope and good adaptation. It can be used to determine the sulfur or chlorine in liquid, gas or solid
materials.
4. Self-developed operation and test software. It will complete the data collecting, processing, saving and printing
automatically.
5. It needs less sample. Only 10μl for each test. The testing time is short. Only 1-2 minutes for each sample.
II. Technical specifications
1. Current: The maximum is ±2mA
2. Output voltage of amplifier: The maximum is ±30V
3. Bias voltage range: (0~500)mv, adjustable
4. Measuring range: 0.3mg/L~10000mg/L(dilutable for high concentration)
5. Repeatability error: ⑴ ≤50% when sample concentration <1.0mg/L
⑵ ≤10% when 1.0mg/L≤sample concentration ≤10mg/L
⑶ ≤5% when sample concentration >10mg/L
6. Temperature control: Ambient to 1000℃, ±1℃
7. Power supply: AC 220V±10V,50Hz±0.5Hz
8. Maximum power consumption: 3000W
9. Ambient temperature: (10~40)℃。
10. Relative humidity: ≤85%
11. Dimension: 700 mm×480 mm×540mm(PC is not included)
12. Net weight: 46kg(PC is not included)
III. Air source
1. Oxygen: 99.5%(An oxygen bottle with a decrement gauge.)
2. Nitrogen: ≥ 99.5%(An nitrogen bottle with a decrement gauge)
I,SYD-0253B Coulometric Chlorine Analyzer Accessories and Documents List
No. | Name | Unit | Qty | Memo |
1 | SYD-0253B Coulometric Chlorine Analyzer | Set | 1 | |
2 | Liquid sampler injector | Set | 1 | |
3 | Microsyringe | Piece | 2 | |
4 | Standard sample | Piece | 10 | |
5 | Silicone rubber pad | Piece | 10 | |
6 | Silicone rubber tube | Centimeter | 20 | |
7 | Pneumatic tube(Ф4) | Meter | 10 | |
8 | Thermocouple | Piece | 2 | |
9 | Power line | Piece | 2 | |
10 | Communication cable | Piece | 1 | |
11 | Quartz tube | Piece | 1 | |
12 | Electrolytic cell | Piece | 1 | |
13 | Quartz cracking tube exclusive clip | Piece | 1 | |
14 | Lenovo computer | Set | 1 | |
15 | Printer | Set | 1 | |
16 | Stirrer | Piece | 1 | |
17 | Operation manual | Piece | 1 | |
18 | Quality certificate | Piece | 1 | |
19 | Product Warranty Card | Piece | 1 |
No. | Name | Specification | Quantity | Memo |
1 | Oxygen | Purity 99.5% | 1 bottle | With pressure regulating valve |
2 | Nitrogen | Purity 99.5% | 1 bottle | With pressure regulating valve |
3 | Glacial acetic acid(HAC) | Excelent pure | 2 boxes |
|
4 | Purified water | Deionized water or redistilled water |