Full Spectrum Direct Reading Spectrometer

1.Product Description TR-TC-9000 Full Direct Reading Spectrometer which adopts the international standards of design and manufacturing technology follows the steps of international spectrum technology and apllies the new all digital techn...

Standard: model: TR-TC-9000 E-mail: robin@pro-trueview.com

1. Product Description
TR-TC-9000 Full Direct Reading Spectrometer, which adopts the international standards of design and manufacturing technology, follows the steps of international spectrum technology and apllies the new all digital technology instead of PMT simulation technology. In addition, in order to make this apparatus' best capability, long term stability and repeability, and the lowest sense lines. The design of vacuum optics room and the technology of all digital excitation light source, advanced CCD detector, high-speed data reading system are applied. This machine is available in metal manufacture and processing industries. It's also one of   the main testing machine in metal metallurgy. Such as quality monitoring, material grade identification, material research and expoding.

2. Main Features
Optimized vacuum optics system
(1).Integrated optics room and Paschen Runge contruction design, making all the spectrum lines focused on the gratings.
(2).Direct-jet type optics technology and MgF2 material lens to make sure ultraviolet wave's best energy of elements, such as C,S,P and N.

All digital excited light source
Adopting all digital and HEPS technology to improve the measuring precision and similarity of sample, which the excetation parameter is setted variously. The samples can show its recurrence by high speed excitation and stable spark.

Functional design excitation equipment
(1).Integrated gas lines, Jet type electrode technology and opened copper spark platform not only can insure the samples' measuring precision but also cater to different shapes' samples.(Including wire materials)
(2).Single plate lens equipment design is easy to user for washing and maintaining the excitation platform and lens.

Highly integrated and direct reading system
(1).Using the 3648 pixels high performance CCD solid testing technology, the spectrum lines in the wave range can  received wholly.
(2).FPGA and high speed data corresponding technology make the data reading function powerfully and progress more quickly.

Visual and easy-operate analysis software
(1).The multi-language CCD full spectrum analysis software, which based on the WINDOWS system, provides powerful data processing ability and testing report output capacity for users under the whole management and controlment of the measuring process.
(2).The apparatus equips many factory calibration curves and many materials to the more advanced solution methods.
(3).The apparatus realizes the full spectrum analysis of intelligent jammer, dark current, the algorithm of appearing and disappearing faces, highly improving the analysis ability of testing machine.
 
Optical system
A、Ptrum optical system of Palmer type - runge structure
B、Aimum wavelength range: 130 ~ 800 nm
C、Liple high performance CCD detector
C、Tgrated optical chamber process molding, resistant to change of    environment temperature
E、Vum light chamber design guarantees C, S, P, could achieve the best  performance.
F、Light room temperature control, constant temperature is 35 °C
G、Rected type optical technology and lens MgF2 materials ensure that the best energy of UV wavelength of C, S, P. 
            
     (Old Style:)                                                                             (New Style)

Distinction
Ole Style:  CCD Technology
New Style:  CCD or CMOS full spectrum analysis technology.
 
Specification and technical parameter:
Item Indicator
Detection substrate Fe  (As a example, other element such as Cu, Al ,Ni,Zn,Sn,Pb and so on)
Detection time According to the type of test sample, usually it is about 25s.
Optical system Palmer type - runge Rowland circle full spectrum vacuum type optical system
Wavelength range 130 ~ 800nm
Focal length 400mm
Detector high-performance CCD array
Electrode Tungsten material spray electrode
Analytical gap Sample Stage Analytical gap:3.4mm
Operating temperature (10 ~ 35)°C
Storage temperature (0 ~ 45)°C
Operating humidity 20% ~ 80%
Argon purity requirements 99.999%
Argon gas inlet pressure 0.5MPa
Argon flow Stimulating flow is about 3.5L/min, maintaining flow is about 0.4L/min.
Stimulating max power 400VA
Light source type brand new adjustable digital light source, high energy preliminary combustion science (HEPS)
Discharge frequency 100-1000 Hz
Discharge current Max 400A
Spark Stand aperture 13mm
Vacuum system Vacuum software automatically control and monitor
Working power 220V AC, 50/60Hz,  the single-phase power supply of protective grounding
Instrument dimensions 860*720*500
Instrument weight 80kg
 
Analyzer The Conventional Range : (Fe Base)