Product Description I.The host can independently accomplish photometric measurement and quantitative measurement,spectral scanning, dynamics,DNA/protein test,wavelength testing and data printing function,ect. II.Using the latest ARM system,...
Product Description
I.The host can independently accomplish photometric measurement and quantitative measurement,spectral scanning, dynamics,DNA/protein test,wavelength testing and data printing function,ect.
II.Using the latest ARM system,built-in double operating system in both Chinese and English,can through the USB interface transfer data storage.
Features
I.Adopt 1200 bar/mm high-performance grating.New light control system makes the instrument light switch more quickly.
II.The instrument adopts 320*240 lattice high-lighted LCD display,direct display scan map,the screen interface simple, convenient use.
III.Socket xenon lamp and tungsten lamp design,after change light no need optical debuggging.
IV.Large sample chamber,Can accommodate 5-100mm various specifications of the cuvette.
V.Can be directly connected to the printer to print patterns and the experimental data.
VI.Optional disk automatic 8 joint cuvettes holder.
Model: | TS-V1800 |
Optical system: | Czerny-turner CT grating monochromator, 1200/mm |
Wavelength range: | 320-1100 nm |
Wavelength accuracy: | 0.3 nm |
Wavelength repeatability: | <=0.2nm |
Measurement range: | -03A-3A |
Photometric accuracy: | <= 0.5%T |
Photometric repeatability: | <= 0.3%T |
Spectral bandwidth: | 2 nm |
Zeroing: | Automatic |
Stray light: | <= 0.05%T |
Stability: | +/- 0.001 A/h @500 nm) |
Display modes: | 320*240 lattice high-lighted LCD display |
Scanning speed: | Fast,Mid,Slow |
Data output (print): | USB Export |
Light source: | Tungsten Lamp ( 6V10W ) |
Input voltage: | AC 220V/50Hz or AC110V/60Hz |