Product Description I.The host can independently accomplish photometric measurement and quantitative measurement, spectral scanning, dynamics, DNA/protein test, wavelength testing and data printing function,etc. II.Using the latest ARM syst...
Product Description
I. The host can independently accomplish photometric measurement and quantitative measurement, spectral scanning,
dynamics, DNA/protein test, wavelength testing and data printing function,etc.
II. Using the latest ARM system, built-in double operating system in both Chinese and English, can through the USB
interface transfer data storage.
Features
I. Adopt 1200 bar/mm high-performance grating. New new light control system makes the instrument light switch more
quickly.
II. The instrument adopts 320 * 240 lattice highlighted LCD display, direct display scan map, the screen interface simple,
convenient use.
III. Socket xenon lamp and tungsten lamp design, after change light no need optical debugging
IV. Large sample chamber,Can accommodate 5-100mm various specifications of the cuvette
V. Can be directly connected to the printer to print patterns and the experimental data.
VI. Optional disk automatic 8 joint cuvettes holder.
Technical parameters
Model: | TS-UV1801S |
Wavelenggth range: | 190~1100nm |
Spectral bandwidth: | (0.5nm,1nm,2nm,4nm,5nm) Adjustable |
Wavelenggth accuracy: | 0.3 nm |
Wavelenggth repeatability: | <= 0.1 nm |
Photometric accuracy: | 0.3%T |
Photometric rang: | -0.3A-3A |
Stray light: | <= 0.05%T |
Stability: | +/- 0.001A/h @500nm |
Baseline flatness: | +/- 0.001A |
Wavelength and the setting mode: | Arbitrarily set |
Scanning speed: | Fast, Mid, Slow |
Wavelength setting method: | automatically |
Detector: | Imported silicon photo diode |
Light source: | Imported long-life tungsten lamp and Deuterium Lamp, |
Display: | 320*240 LCD |
Data output: | USB Export |
Printer : | Optional |
Analysis Software: | Standard accessories |
Input power: | AC220/50Hz or AC 110/60Hz |
Weight: | 25kg |