Petroleum Instruments

Distillation Apparatus(Low Temperature Double Tubes))

model: SYD-6536B
Standard: ASTM -- D86
E-mail: robin@pro-trueview.com


Summary
  The instrument is designed and made as per standards GB/T6536 Test Method for Distillation Characteristics of Petroleum Products and ASTM D86 Standard Test Method of Distillation for Petroleum Products at Atmospheric Pressure. It is suitable to determine the distillation characteristics of gasoline, aviation gasoline, jet fuels, solvent having special boiling point, naphtha, diesel oil, distillate fuels, and other petroleum products.
  This instrument is a kind of double-units instrument which the minimum temperature of the condenser pipe can reach 0℃.It is an ideal distillation tester for petroleum products under low temperature condition.
 
I. Main technical features
1.The instrument adopts low temperature and double units working mode. The minimum temperature can set to 0℃.It can
   do test for one sample or two samples at same time.It highly improves the work efficiency.
2.It adopts digital temperature controller, stainless-steel heater, new type refrigeration compressor and electric stirrer to
   constitute constant temperature bath. The temperature in the bath is uniform,can fully meet the requirements of test.
3.It adopts special designed structure, cooling speed is fast.The bath temperature can cool down from room temperature
   to 0℃ within 50 minutes.
4.It adopts special heating furnace to ensure the safety of test. Heating power can be continuously adjusted to meet the
   requirements of test.
 
II. Main technical specifications
1.Power supply:  AC(220±10%)V, 50Hz
2.Heating power: 1000W×2.
3.Receiving cylinder: 100ml, division value is 1ml.
4.Distillation flask: 125ml,meet the requirements of GB/T 6536 and ASTM D86
5.Thermometer:(-2~300)℃ and (-2~400)℃. Division value 1℃.
6.Flask support board: SiC, bore diameter is: φ32mm, φ38mm, φ50mm.
7.Temperature controller:
1) Range:  0℃~60℃   2) Accuracy:  ±0.5℃    3) Display:  LED
    Note: The temperature control is to control the temperature of condensate in the condenser pipe.
8.Ambient temperature:  ≤30℃
9.Relative humidity:  ≤85%
10.Cooling system: New type refrigeration compressor
11.Maximum power consumption:  4000W
12.Dimension: 730mm×530mm×580mm
Return

Shanghai Trueview Scientific Instrument Co. LTD
Phone: 0086-18007322239 (Robin)
Fax: 0086-734-8770418
Cell: 0086-15367067787
Wechat: 0086-18007322239
Email: robin@pro-trueview.com
Add: Room 118, Block 21, No.1-42, Lane 83 ,Hongxiang North Road, Wanxiang Town, Pudong New Area, Shanghai City, China
Copyright © 2018.Trueview Scientific All rights reserved.