Petroleum Instruments

Ultraviolet Fluorescence Sulfur,Nitrogen-in-Oil Analyzer

model: SYD-0689N
Standard:
E-mail: robin@pro-trueview.com


Summary 
  The instrument is designed and manufactured according to the standards SH/T 0689-2000 Standard test method for determination of total sulfur in light hydrocarbons,motor fuels and oils by ultraviolet fluorescence,SH/T 0657-2007 Standard test method for trace nitrogen in liquid petroleum hydrocarbons by syringe/inlet oxidative combustion and chemiluminescence detection and ASTM D 5453 -2006 standard.It is widely used in petroleum,chemical industry,electric power,coal,food,environment proection and others.At present,SYD-0689N is the advanced instrument to analyze the sulfur,nitrogen in China.

I.Main technical characteristics
1.    The instrument can detect the both sulfur and nitrogen,also could test the solid sample or liquid sample.
2.    With the desktop computer,Windows system,man-machine direct conversation platform,It can set all prameters and
       make the test by the keyboard operation.
3.    It adopts the control technology of the computer to automatically collect,process,store and print all data.with the wide  
       linearity range of measurement,strong capacity of resisting disturbance,easy operation and reliable test results.
4.    With the imported fluorescence excitation source,membrane desiccator,optical filter,metal encapsulated
       photomultiplier,It has the features of the high sensitivity,fast,stability,high precision,good consistency.
5.    With the advanced temperature control system,fast and steadily warming rate,long service life,the temperature control
       accuracy ±3℃ during the temperature range,It can meet the test requirements.  

II. Main technical specifications
1.    Determination method: Ultraviolet fluorescence method to detect sulfur,chemiluminescence to detect nitrogen
2.    Sample types:Liquid,solid and gas
3.    Sample injection quantity:Solid:1-20mg; Liquid:5-20μL; Gas:1-25mL
4.    Measurement range:  (0.2~8000)mg/L(High concentration can be diluted)
5.    Measurement accuracy:
Concentration values (ppm) Injection quantity(µL)  RSD(%)
0.2 20 25
5 10 10
50 10 5
100 10 3
5000 10 3
6.    Temperature control range: Ambient to 1050℃,accuracy: ±3℃
7.    Air supply requirements:High purity argon:above 99.995% ; 

                                                High purity oxygen:above 99.9%                      
Note:The water content of the gas must be less than 5ppm
8.    Environment temperature:Less 30℃(The laboratory with air condition)
9.    Relative humidity: ≤85%
10.    Power supply: AC220V±22V,50Hz±0.5Hz,more than 2000 W
11.    Dimension:Host:305(L)×460(W)×440(H)mm;                                                
        Temp controller:550(L)×460(W)×440(H)mm,No computer
12.    Net weight:Host::15kg, Temp controller:30kg
 


Return

Shanghai Trueview Scientific Instrument Co. LTD
Phone: 0086-18007322239 (Robin)
Fax: 0086-734-8770418
Cell: 0086-15367067787
Wechat: 0086-18007322239
Email: robin@pro-trueview.com
Add: Room 118, Block 21, No.1-42, Lane 83 ,Hongxiang North Road, Wanxiang Town, Pudong New Area, Shanghai City, China
Copyright © 2018.Trueview Scientific All rights reserved.