Total Acid Number Tester(Potentiometric Titration)
The instrument uses potentiometric titration principle. Recording the titration of the electrode potential and the titration volume, find out the equivalence point and corresponding titrant volume, and thus obtained the sample pH value. This instrument can accurately detect the acid value of transformer oil, turbine oil, anti oil, diesel, gasoline and other petroleum products;Widely used in chemical industry, electric power, oil and other industries, it’s the ideal products to current meter acid number of petroleum products. Standard:GB/T264 Determination of acid number、GB7599-87 Determination of acid number in transformer and turbine oils in service(BTB)、GB/T7304-2000. GB/T 18609-2001. D664—01.ASTM D664-2011)
I. Main technical features
1. Windows operating platform, Users can dialogue with the computer directly. It’s easy to operate. Has a workstation
features.
2. Titration curve real-time display, the titration curve and data storage and printing.
3. Imported titration unit.Measuring accuracy better stability.
4. Automatic cleaning, automatic constant liquid.
5.Automatic detection of end points, Automatic filtering false end, At the same time can determine the endpoint of artificial
selection.
6. With continued dosing function.
II. Main technical specifications
1.Acid number measurement range: ≥0.05 mgKOH/g
2.Potential measurement range: (0~1800.0)mV
3.Burette volume: 10mL
4.Minimum titration volume: 0.001mL
5.Burette precision: ±0.1%F•S
6.Precision: Fit or above the relative standards
7.Power supply: AC(220±11)V ,50±1Hz
8.Environment temperature: 5℃~35℃
9.Relative humidity: Less 80%
10.Outline dimension:About 350㎜×280㎜×178㎜(L×W×H、Without PC)
11.Net weight: Main unit 14Kg