Gas Chromatography
model: TS6891N
Standard:
E-mail: robin@pro-trueview.com
Description:
1.1 Injector flow, column carrier flow, detector gas flow are controlled by EPC.
1.1.1 Micro-channel gas integrated board
1.1.2 Imported micro-ratio and flow/pressure sensor
1.1.3 Temperature self-compensation to ensure the accurate flow, its pressure controlling can precise to 0.01Psi.
1.2 Industry-leading detector design.
1.2.1 All the detectors with high-precision electronic flow/pressure control.
1.2.2 It can be matched with packed column and capillary column.
1.2.3 The amplified board after special ceramics treatment can keep stable in wet season.
1.3 It is safe and reliable with the functions of self-ignition and hydrogen leakage self-protection.
1.4 Accurate tracking design for furnace temperature.
1.5 Accurate programmed heating control for the furnace temperature.
1.5.1 Inside temperature controlling precision: <0.01
1.5.2 Temperature between inside points: ≤1℃
1.5.3 Quick heating and cooling, the max programmed heating speed: 120℃/min, balance time: 10 sec, cooling time: 300℃-50℃ for 6 minutes.
1.5.4 The temperature precision is absolutely controlled(compared with the real temperature, the difference is less than 0.3℃).
1.6 The achievement of gas circuit cutting technology
1.6.1 Complex samples separation can be analyzed by multidimensional chromatograph.
1.6.2 It saves the analysis time and period by blowback function of the chromatograph column.
1.7 Chromatograph columns can be heated individually, and the using temperature can be set according to the different columns, so that make every column separately.
1.8 TS6891N adopts the advanced power distribution management shunt. One 3000-watt UPS can provide stable working power for 10 sets of TS with the normal power of 2250w/set.
1.9 TS6891N adopts the imported valve configuration system, which can carry 8 valves (4-way valve, 6-way valve, 10-way valve or liquid sampling valve) to automatically control and operate by sequence.
1.10 TS6891N uses 8-digit auto-injectors (you can choose Agilent 7683/7693 liquid auto-injector).
Auxiliary facilities for option
DK-400A: a new generation of automatic headspace injector.
Application fields
TS6891N is widely used for the analysis of oil extraction and refining, petrochemical, fine chemical, food safety, environmental protection, electronic high-purity gas industry, scientific research institutes, QC, safety supervision, public security forensic and national security and so on.
Specifications:
Column oven | |
Furnace dimensions | 28 x 30 x 18cm |
Temperature controlling rang | Higher than the room temperature of 5℃-450℃ liquid nitrogen cold trap: -80℃-400℃,dry ice cold trap:-55℃ to 400℃ |
Temperature setting precision | 0.1℃ |
Max programmed heating speed | 120℃/min |
The longest method run time | 999.99 min |
The highest step of program | 7 steps |
Programmed heating rate | 0.1-120℃/min(increment of 0.1℃) |
Column loss can be compensated (dual channel column) | |
Heating area | |
Six individual controlling heating areas (do not content furnace temperature control two injectors two detectors and two auxiliary heating areas ). | |
The maximum using temperature in auxiliary heating areas | 300℃ |
Detectors | |
FID | |
Maximum using temperature | 450℃ |
Minimum LOD: | ≤2.5pgC/s[n-C16] |
Linear dynamic range | 107(±10%) |
Data collection frequency | Max 100 Hz |
Adapted to packed columns and capillary columns | |
TCD | |
Maximum using temperature | 400℃ |
Minimum LOD | <400 pgC3H8/ml(He) |
Linear dynamic range | 105(±5%) |
Data collection frequency | Max 100 Hz |
The design of monofilament TCD and micro-ECD adapted to packed columns and capillary columns | |
FPD | |
Maximum using temperature | 250℃ |
Minimum LOD | <4 pg S/sec<1 pg P/sec |
Linear dynamic range | >103S,>104P |
Optional | 105gS/gC,106gP/gC |
Adapted to packed columns and capillary columns | |
ECD | |
Maximum using temperature | 400℃ |
Type of gas compensation detector | 5% methane/argon/nitrogen |
Minimum LOD | <0.01 Pg/sec lindane |
Linear dynamic range | >5 x 105 |
Data collection frequency | Maximum 100 Hz |
Radioactive source | <12 mCi 63Ni |
Adapted to packed columns and capillary columns | |
NPD | |
Maximum using temperature | 450℃ |
Minimum LOD | <0.2 pg N/sec,<0.2 pg P/sec |
Minimum LOQ | <3 pg C/sec |
Linear dynamic range | 105 N,105 P |
Adapted to packed columns and capillary columns |