Gradient High Performance Liquid Chromatography
model: HPLC-600B
Standard:
E-mail: robin@pro-trueview.com
Description:
1) Hardware has VP function, can record maintenance information and operation which can meet the requirement for the GLP/GMP.
2) Humanized designed for the system control, which has the function of clock, thermometer, hygrometer and so on.
3) Multi software, can meet the requirements for different customers. Chromatogram data processing system is embedded in detector, has VI5890, N2000, N3000 for chosen.
4) Simple system for customer to skill for the software operation,
5) High performance. Using stronger Teflon gasket ring, floating piston setting way, to enhance the stability and durability for the instrument.
6) Detector use the imported deuterium lamp, light battery, dual-beam monochromator which consist of 1200 unit/mm concave raster, precision processed double lens flow cell, high precision micro-processor for adjusting the wavelength, two-way high-speed sampling rate, to ensure low noise, low drift and high sensitivity.
7) Even mixing mixer. It has good repeatability for the baseline when using gradient program acquisition.
8) Elegant surface to meet the visual sight requirement for customer.
Application Area
scientific research, medicine detection, food detection, chemical analysis, environmental detection.
Specifications:
HPLC600C High-Pressure Constant-Current Infusion Pump |
|
Infusion way |
Micro Volume Series Dual-Piston |
Infusion Pressure |
0-9999Psi |
Flow Rate Erro | ≤0.5% |
Flow Rate Range | 0.001-9.999mL/min(in 0.001mL/min Increments) |
Stability Error for the Flow Rate | ≤0.2%RSD |
Gradient Error | ±1%(0-100% Water /Acetone Solvent) |
Pressure Pulsation | <15Psi(Flow Rate1mL/min, Pressure 600-1600Psi) |
Air Tightness | Pressure 5400Psi,Time 10min,Pressure Drop<400Psi |
Time Program Function | Have |
Size | 260mm x 130mm x 420mm(W*H*D) |
Weight | 11kg |
Operation Temperature | 4-40℃ |
SPD600C UV Detector |
|
Wavelength Range | 190-700nm |
Wavelength Error | ≤1nm |
Wavelength Repeatability Error | ≤±0.1nm |
Dynamic Noise | ≤±0.75 x 10-5AU(Methyl Alcohol,1ml/min,254nm,20℃) |
Static Noise | ≤±0.5 x 10-5A(Empty Cell, Response Time 1 second 20℃) |
Dynamic Baseline Drift | ≤±1 x 10-4AU/h(Methyl Alcohol,1ml/min,254nm,20℃) |
Static Baseline Drift | ≤0.5 x 10-4(Empty Cell, Response Time 1second, 20℃) |
Linear Range | ≥104 |
Limit of Detection | ≤1 x 10-9g/mL(Naphthalene Methyl Alcohol Solvent) |
Qualitative Repeatability | RSD6≤0.1% |
Quantitative Repeatability | RSD6≤0.5% |
Spectrum Band Width | 6nm |
Flow Cell Volume | 8μL |
optical path | 10mm |
Time Program Function | Have |
Size | 260mm x 130mm x D420mm(W*H*D) |
Weight | 11kg |
Operation Temperature | 4-40℃ |