Gas Chromatographs
model: GC-40A(03A)
Standard:
E-mail: robin@pro-trueview.com
Introduction:
1. GC-4000A series GCs are suitable for application in analysis of gases,liqids,and mixture of gases and liquids.
2 . Our products are widely uesd in the field such as petroleum, chemical industry, coal mining,food,enviromental sciences,pesticide, hygiene and epidemic prevention,quality inspection etc.
3. 5 detectors to choose. Thermal conductivity Detector: Uesd for macro analysis of gases and liquids
Flame ionization Detector: Used for trace lever analysis of sulfide and phosphide.
4 .Modular design,the big sized columns oven makes installation and maintenance convenient.Up to 3
chromatogram columns can be installed in an oven. Possible to install any 3 detectors and 3 independent circuit systems and 3 signal outputs.
5 .Fiexible packed columns and capilillary columns system.
6 .Temperature Rising Program
Number of programmable Ramp: 10
Ramp Speed: 0~40 ℃/min (min increment 0.1 ℃)
7 .Sample injection
Capillary Column.Split/Splitless,packed Column
Cold Headed column
8 .Automatic back door control.Oven-heating protection
Features
● Five detectors to choose from: thermal conductivity detector (TCD), flame ionization detector (FID), flame photometric detector (FPD), nitrogen phosphorus detector (NPD), electron capture detector (ECD).
● Features a large column oven, which allows up to three columns to be installed simultaneously.
● Can be fitted with up to three detectors of any type. In addition, can be fitted with three independent signal processing circuits that provide three output signals.
● Flexible packed and capillary column interface. Capillary columns of many different widths can be installed.
● Features a stable temperature control system, with 10 segment programmable temperature control, and over-heating alarm and protection.
● Automated back door control helps speed up column oven cooling.
● Up to three pressure controlled gas paths are available. As well as digital display of pressure and flow. Pressure flow has a display resolution of 0.1 psi.
● Additional available devices: blowback device, pre-cutting device, methane conversion furnace, cracking furnace
Main Parameters:
FID | TCD | |
Sensitivity |
S≥8000mv mL/mg (benzene) |
|
Limit of detection |
M≤1×10-11 g/s (hexadecane) |
|
Noise | ≤5×10-14A | ≤0.01mv |
Scope of Linearity | 107 | 104 |
Drift | 1×10-13 A/30min | 0.02mv/30min |