Particle Size Analyzer

Intelligent Broad Size Laser Particle Size Analyzer

model: LPS-T28A
Standard:
E-mail: robin@pro-trueview.com


Brief introduction:
      LPS-T28A full automatic wet laser particle size analyzer adopt MIE scattering principle, measure size is from 0.01 to 2000 micron, Which offer reliable and repeatable particle size analysis for a diverse range of applications.It use dual-beam& multiple spectral detection systems and side light scatter test technology to significantly improve precision and performance of test, on behalf of the domestic advanced level in the field.

Main Specifications:
Model Name LPS-T28A LPS-T28D
Standard ISO13320-1:1999, GB/T19077.1-2008, Q/JWN001-2009
Principle MIE scattering principle
Measuring Range 0.01-2000μm 0.01-1200μm
Channels Number 127 127
Accuracy    <1%   (Deviation of  D50 on national standard sample)
Repeatability <1%   (Deviation of  D50 on national standard sample )
Light source High performance He-Ne Laser (λ= 632.8nm,  P>2MW)
Lifetime>25000hours
 
Dispersion
 Method
Ultrasonic Frequency:40KHz   Power:35W,   Time: ≥1S
Stir Revolutions Speed: 0-300RPM (Adjustable)
Circulate Rated Flow:8L/min    Rated Power:10W
Sample Pool Volume:350mL
Micro-
Sample Pool
Volume: 10mL ( Available)
Operation Mode Full automatic/ manual control, freely choose
Optical bench alignment system Full automatic,precision is up to 0.1um
Software function Analysis mode Free Distribution, R-R Distribution, Logarithm Normal Distribution, Mesh number classification etc.
Statistic Method Volume Distribution, Quantity Distribution
 
Statistic Comparison Several Testing Results of samples
Different batches of samples testing result,
Samples before and after processing,
Test result of samples in different time.
 
User-defined Analysis Figure out percentage according to the particle size
Figure out particle size according to the percentage
Figure out percentage according to the particle size range
Meet demands of representation of particle test in different industries
Test Report Word, Excel,Photo( Bmp), Text etc
Multiple-language Support Multiple language Support
Intelligent operation  Automatically control water inflow, dispersion,test and analysis.Better Repeatability after remove human-factor
Testing speed <2min/time
Outer dimension L66cm*W32cm*H40cm
Net weight 65Kg
 
Main Features:
1)TrueviewPatent -Optical path design

Converging light Fourier transform light patented technology, make scattering light not be restricted to lens aperture. Double spectrum design make its precision achieves from 0.01um to 2000μm,and Dual-laser orthogonal light make use of the semiconductor auxiliary laser extend the test angle from 45 degree to 135 degree, ensure receive all the angles of signals. 
2) Expand Size Range
LPS-T28A measurement range significantly expanded to 0.01um, particularly suitable for small samples of the user, the sub micron particle test results reached the international level 
3)Full built-in dispersion system
The ultrasonic dispersion, mechanical stirring and circulation channel reasonable integrated in the interior of the instruments, ensure  the uniform dispersion and distribution of particles in the testing process, effectively avoid uneven distribution, large particle deposition phenomenon caused by long pipe line of outside dispersion system , to ensure the accuracy of the test results. 
4)Intelligent Operation mode
With intelligent automatic mode of operation, to achieve a key test, as long as according to the prompt addition of sample, click the "test", all processes including water-supply, dispersion, circulation, testing, cleaning, data record, data analysis, save and print are automatically completed, not only reduce the testing workload, but also eliminate the interference of human factors, to further improve the accuracy and authenticity of testing results.
5)Automatic Optical path alignment system
The precision of four hybrid stepping motor in the automatic system of optical components, micro precision of 0.1um, the instrument of light path is always at its best to eliminate manually on the light path and the troubles and difficulties but also enhance the accuracy and stability of test results.
6)Authoritative calibration method
In the entire test range using the national standard materials for calibration, only a year calibration once the instrument can. Calibration method is simple and quick.
7) Unconstrained free fitting technique
Trueview original unconstrained free fitting technology, particle size analysis is not affected by any functional constraints, can be a true reflection of the distribution of particles.
 
Instrument principle diagram

 
Adopt Patents Technology:
1.Optical bench design is protected by patent No.- ZL 2014 2 0378380.8, 
2.Optical bench alignment system is protected by patent No.- ZL 2013 2 0835882.4.
3.Mie scattering principle application patent No.- ZL 2013 2 0812021.4.
4.Dual laser beam orthogonal application is protected by patent No.-ZL 2007 2 0025702.0
 
Instrument Test operation interface:
    After the background test, click on Energy spectrum test” in "test view". And the system will display "test view". Observe the Spectral curve and concentration, After the test result becomes stable, click the “save the results”. The System will automatically save the test result at the preset time intervals
 
                                                                               Energy Spectrum Test View
 
Test Report and Its Description:
 
1) D10 : Particle diameter, < 10% particle volume summation percent of total particle’s volume
2) D50 : Particle diameter, < 10% particle volume summation percent of total particle’s volume
3) D90 : Particle diameter, < 10% particle volume summation percent of total particle’s volume
4) DAV:The average particle size of the particle swarm
5) S/V: Volume specific surface area; surface area per unit volume of particles
        
 Particle size analysis diagram:    
1)Horizontal axis is the size of the particle, which is the logarithm distribution.
2)Left column is the volume of the cumulative percentage, corresponding to the upward trend curve
3)Right column is the volume percentage of a certain interval, corresponding to the histogram or the ups and downs of the curve.
4)The test result of data list is corresponding to the analysis chart.
 
Application:

LPS-T28 is widely used in mineral materials, cement, ceramics, chemicals, paint, emulsion, dyes, pigments, fillers, chemical products, catalysts, drilling mud, abrasive, lubricant, coal, sand, dust, bacteria, cells, food, additives, pesticides, explosives, graphite, photosensitive materials, fuel, and ink metal and non metal powder, calcium carbonate, kaolin, coal water slurry and other powder material.

Return

Shanghai Trueview Scientific Instrument Co. LTD
Phone: 0086-18007322239 (Robin)
Fax: 0086-734-8770418
Cell: 0086-15367067787
Wechat: 0086-18007322239
Email: robin@pro-trueview.com
Add: Room 118, Block 21, No.1-42, Lane 83 ,Hongxiang North Road, Wanxiang Town, Pudong New Area, Shanghai City, China
Copyright © 2018.Trueview Scientific All rights reserved.